Breathing Circuit Leak Caused by Breathing Tube Holder: A Case Report

Keywords : Anesthesia, Breathing circuit, Circuit leak, Tube holder

DOI: 10.5005/jp-journals-10045-00201

License: CC BY-NC 4.0

Published Online: 29-03-2022

Copyright Statement:  Copyright © 2021; The Author(s).


Abstract

Breathing circuit leak, obstruction, or misconnects contribute to around 20% of the critical incidents occurring during anesthesia which if not immediately identified may result in serious harm. Circuit damage can occur due to extra-manipulation of the circuit after fixing it inside the tube holder or mishandling of the tube holder with the circuit in situ. Damage to the breathing circuit can be prevented by making the edge of the tube holder groove blunt or by adding some cushion material over the edge of the groove. The anesthesiologist must be vigilant about circuit leaks while administrating anesthesia.


HTML PDF Share
  1. Cooper JB, Newbower RS, Long CD, et al. Preventable anesthesia mishaps: a study of human factors. Anesthesiology 1978;49(6):399–406. DOI: 10.1097/00000542-197812000-00004
  2. Caplan RA, Vistica MF, Posner KL, et al. Adverse anesthetic outcomes arising from gas delivery equipment: a closed claims analysis. Anesthesiology 1997;84(7):741–748. DOI: 10.1097/00000542-199710000-00006
  3. Miller RD. Miller's anesthesia. 7th ed. Philadelphia: Elsevier Churchill Livingstone; 2010. p. 669.
  4. Doo AR, Lee SK. Anesthesia machine breathing tube holder. Korean J Anesthesiol 2015;68(1):87–88. DOI: 10.4097/kjae.2015.68.1.87
PDF Share