Twisting of Tail End of the Reservoir Bag Malingering as Stuck Valve during Positive-pressure Ventilation

Poonam Kumari, Amarjeet Kumar, Chandni Sinha, Ajeet Kumar, Rajesh Kumar

Keywords : Airway obstruction, Pediatric, Reservoir bag

DOI: 10.5005/jp-journals-10045-00155

License: CC BY-NC 4.0

Published Online: 29-09-2021

Copyright Statement:  Copyright © 2020; The Author(s).


Abstract

The most commonly used open circuit is The Jackson Rees modification of the Ayre's T-piece (Mapleson-F system) in pediatric patients because it has low resistance and nominal dead space. Here, we report a case in which we used the Jackson Rees circuit with bag tail valve for ventilation in pediatric patient weight 12 kg. During positive-pressure ventilation, we felt resistance and the patient was not ventilating. High pressure in the bag was being formed despite the valve was fully open. On inspection, we found out the cause was a twisted bag tail end as shown in Figure 1. We straighten the tail end of the bag, which aided in releasing air pressure and the patient started getting ventilation. The obstruction of the expiratory limb of the reservoir bag may be due to a stuck valve or due to twisting of its tail end. Due to high flexibility, twisting of the tail end of reservoir bag commonly found when it was held with tail end up position. Due to the twisting of the tail end of the reservoir bag, the expiratory limb becomes closed and high flow oxygen started going to the inspiratory limb. This may result in CO2 retention and barotraumas of the lungs if the obstruction was not released. We can prevent twisting of the tail end of the bag by keeping the Jackson Rees circuit in a vertical position with the tail end facing down.


HTML PDF Share
  1. Jerwood DC, Jones SE. HME filter and Ayre's t piece. Anaesthesia 1995;50(10):915–916. DOI: 10.1111/j.1365-2044.1995.tb05873.x.
  2. Mapleson WW. The elimination of rebreathing in various semiclosed anaesthetic systems. Br J Anaesth 1954;26(5):323–332. DOI: 10.1093/bja/26.5.323.
  3. Wills BA, Pender JW, Mapleson WW. Rebreathing in a T – piece: volunteer and theoretical studies of the Jackson – Rees modification of Ayres’ T - Piece during spontaneous respiration. Br J Anaesth 1975;47(12):1239–1246. DOI: 10.1093/bja/47.12.1239.
  4. Dorsch JA, Dorsch SE. Understanding anaesthesia equipment. Mapleson Breathing System. 5th ed., New Delhi: William and Wilkins; 2008. pp. 209–221.
PDF Share
PDF Share

© Jaypee Brothers Medical Publishers (P) LTD.